Research Article

The Effects of Flattened and Flattening-Filter-Free Beam on Treatment Plans

Yue Yan1, Poonam Yadav2, Michael Bassetti2, MD, PhD, Paul Harari2, MD, Bhudatt R. Paliwal*3

1Department of Medical Physics and Human Oncology, University of Wisconsin-Madison, WI 53792, USA
2Department of Human Oncology, University of Wisconsin-Madison and University of Wisconsin, Riverview Cancer Center, Wisconsin Rapids, WI 54494, USA
3Department of Medical Physics and Human Oncology, University of Wisconsin-Madison, WI 53792, USA

*Corresponding author: Dr. Bhudatt R. Paliwal, University of Wisconsin Madison, K4/314 CSC, 600 Highland Ave Madison, WI 53792, USA, Tel: (608) 263-8514; Email: paliwal@humonc.wisc.edu

Submitted: 09-08-2015 Accepted: 10-06-2015 Published: 10-19-2015

Download PDF

_________________________________________________________________________________________________________________________

 

Article

Abstract

We investigated the dosimetric and biological differences between the flattened and flattening-filter-free (FFF) beam plans for the 6 and 10 MV photon beams from the Varian TrueBeam system. One hundred and four treatment plans were created for 13 cases involving 3 different cancer sites: head & neck, lung and prostate. Significant dose sparing was obtained with FFF beams for the head & neck cases, especially for cases with large field sizes (≈16×20 cm2). Overall, the FFF beams provide similar target coverage as the flattened beams. Additionally, the FFF beams resulted in improved dose sparing to organs at risk compared to the conventional flattened beams, for both static intensity modulated radiotherapy and volumetric modulated arc therapy plans.

Keywords: Treatment Planning; Flattening-Filter-Free; Intensity-Modulated Radiotherapy

Introduction

Flattening filters have been used for decades to provide uniform beam profiles. With the rapid developments of intensity- modulated radiotherapy (IMRT), the flattening filter is no longer an indispensible component of a modern radiotherapy system. CyberKnife [1] and Tomotherapy [2] use unflattened beam to deliver acceptable dose distribution. In recent years, the potential application of the flattening-filter-free (FFF) beam in modern linac system is being explored. Removal of the flattening filter from the path of the photon beam leads to significant changes in several physical characteristics of a conventional beam [3]. Compared to the flattened beams, the  FFF beams have: (1) increased dose rate [4-7], (2) decreasedexternal scatter from the flattening filter [8], (3) reduced neutron contamination for the high energy beam (>10 MV) [9-11], and (4) improved dose calculation accuracy [3]. However, the non-uniform beam profile of the FFF beam requires higher monitor unit (MU). Additionally, the soft spectrum of the FFF beam is likely to increase the dose to organs-at-risk (OARs). Several groups [11-15] have investigated the benefits of FFF beam. Most of them focused on the time efficiency obtained by the FFF beam compared to the flattened beam. The dosimetric and biological differences between the flattened and FFF beams are still unclear. In this work, we evaluated these differences for clinical treatment plans between the flattened and FFF beams.

Materials and Methods

Thirteen patients with three different cancer sites (head and neck (5 patients), lung (4 patients), and prostate (4 patients)) were randomly selected. A case number was used to refer to each plan. TrueBeam STx linac (Varian Medical Systems, Palo Alto, CA) was commissioned on the EclipseTM treatment planning system based on the clinical golden data[16]. Beam energies were 6 and 10 MV. An Anisotropic-Analytical-Algorithm (AAA) was used to calculate the dose distribution for both static IMRT and volumetric modulated arc radiotherapy (VMAT) plans. The dose calculation grid was 2.5 mm for all treatment plans. Beam modalities consisted of flattened and FFF beams. High definition (120 leaves) multi-leaf collimator (MLC) was used to modify the beam fluence for all treatment plans. The widths of the MLC were 2.5 mm and 5 mm in the center and peripheral region, respectively. In order to make a reasonable comparison, all optimization parameters, such as, arc number, beam angle, optimization weight, were kept identical for the flattened and the corresponding FFF beam plans. For the VMAT plans, maximum dose rates of the FFF beam (1400 MU/ min for 6 MV and 2400 MU/min for 10 MV) were used to design the plans. For static IMRT plans (sliding window), the dose rate was 600 MU/min for both flattened and FFF beam plans. For sliding window static IMRT plans, the speed limit of the MLC may impact the dose to OARs for high dose rate. It is more likely for large field sizes (e.g. ). Reducing the dose rate can eliminate the impact of the interplay effect of the MLC to dose distribution. For VMAT plans, however, this problem does not exist because the Eclipse software will automatically select the dose rate it can use to deliver the radiation. For each patient, 8 treatment plans were designed.

An in-house-developed Matlab (Math Works, Natick, MA) code based on the CERR software [17] was used to read the DICOM data extracted from Eclipse to statistically analyze the results. All planning evaluation parameters (e.g. maximum dose, mean dose, and volume) for each organ were benchmarked with the Eclipse treatment planning software. Excellent agreement was obtained for all treatment plans.

Biological metrics was based on the dose-volume-histogram (DVH). It consisted of biological effective dose (BED) [18], equivalent uniform dose (EUD), tumor control probability (TCP), and normal tissue complication probability (NTCP) values. Gay and Niemierko’s model [19,20] was used to perform calculations. Values of biological parameters (α⁄β, TD50, TCD50, a, and γ50) were selected from the published references [19-24]. The biological parameters of OARs are summarized in Table 1.

Table 1. Biological parameters used for OARs to calculate NTCP values.

Radiation Table 21.1


Results

For all cases, the FFF beam provided similar target coverage as the flattened beam. Significant dose sparing to OARs was obtained for the head and neck cases, especially for cases with large field sizes ( ≈16×20 cm2). The statistical analysis results
are shown in Table 2. For each cancer site, in general, the FFF beam provides improved dose sparing compared to the flattened beam. Overall, the FFF beam provides lower mean and maximum dose to OARs, leading to improvement in dose sparing compared to the flattened beam, for both static IMRT and VMAT plans. For critical organs such as brainstem and parotid, the reduction in the mean dose reached up to 14% and 9%, respectively. The reduction in maximum dose was as much as 10% and 5%, respectively. In lung cases, the larynx had the most apparent dose sparing from the FFF beam plans. The mean dose reduction of the FFF beam reached to 8%, that led to decreased NTCP values compared to the flattened beam. For prostate cases, the dose sparing of the FFF beam was not clinically significant for both static IMRT and VMAT plans.

Discussion

In our investigation, the head and neck cases showed the most apparent differences between the flattened and FFF beams in both static IMRT and VMAT plans. Compared to two other cancer sites, some head and neck cases in our study required relative large field sizes (≈16×20 cm2) to cover the target. The noticeable dose sparing of the FFF beam compared to the flattened beam could be explained by the peak beam profiles of the FFF beam.

Table 2. Relative dosimetric and biological ratios (FFF/flattened) for all clinical cases.

Radiation Table 21.2

For small field sizes (e.g. ≈ 6×6 cm2), the differences in the beam profiles between the flattened and the FFF beams are minor. However, for large field sizes (e.g. ≈ 20 ×20 cm2), the FFF beam provides lower dose to the out-of-field region compared to the flattened beam for both 6 and 10 MV beams.

Conclusion

In this work, we investigated the dosimetric and biological differences between the flattened and FFF beams in 13 clinical cases. Significant dose sparing was obtained with FFF beams for head and neck cases, especially for cases with large field sizes (≈16×20 cm2). In general, the FFF beam can provide similar target coverage as the flattened beam with improved dose sparing to OARs.

Acknowledgements

The authors in this work would like to thank Dr. Michael Bassetti in Human Oncology Department, University of Wisconsin Madison, for inspiring discussions on the clinical impact of the improved dose sparing effect of the FFF beam.

References

References

1.Adler Jr. JR., Chang SD, Murphy MJ, Doty J, Geis P et al. The Cyberknife: A Frameless Robotic System for Radiosurgery. Stereotact Funct Neurosurg. 1997, 69:124–128.

2.Mackie TR, Holmes T, Swerdloff S, Reckwerdt P, Deasy JO et al. Tomotherapy: a new concept for the delivery of dynamic conformal radiotherapy. Med Phys. 1993, 20(6):1709- 1719.

3.Georg D, Knöös T, McClean B. Current status and future perspective of flattening filter free photon beams. Med Phys. 2011, 38(3):1280–1293.

4.O’Brien PF, Gillies BA, Schwartz M, Young C, Davey P. Radiosurgery with unflattened 6-MV photon beams. Med Phys. 1991, 18(3):519-521.

5. ">Cashmore J. The characterization of unflattened photon beams from a 6 MV linear accelerator. Phys Med Biol. 2008, 53(7):1933-1946.

6.Kragl G, afWetterstedt S, Knäusl B, Lind M, McCavana P et al. Dosimetric characteristics of 6 and 10MV unflattened photon beams. RadiotherOncol. 2009, 93(1):141-146.

7.Vassiliev ON, Titt U, Pönisch F, Kry SF, Mohan R et al. Dosimetric properties of photon beams from a flattening filter free clinical accelerator. Phys Med Biol. 2006, 51(7):1907- 1917.

8.Cashmore J, Ramtohul M, Ford D. Lowering whole-body radiation doses in pediatric intensity-modulated radiotherapy through the use of unflattened photon beams. Int J Radiat Oncol Biol Phys. 2011, 80(4):1220-1227.

9.Kry SF, Howell RM, Titt U, Salehpour M, Mohan R et al. Energy spectra, sources, and shielding considerations for neutrons generated by a flattening filter-free Clinac. Med Phys. 2008, 35(5):1906-1911.

10.Kry SF, Titt U, Pönisch F, Vassiliev ON, Salehpour M et al. Reduced neutron production through use of a flattening-filter- free accelerator.Int J Radiat Oncol Biol Phys.2007, 68(4):1260-1264.

11.Mesbahi A. A Monte Carlo study on neutron and electron contamination of an unflattened 18-MV photon beam.Appl Radiat Isot. 2009, 67(1):55-60.

12.Thomas EM, Popple RA, Prendergast BM, Clark GM, Dobelbower MC et al. Effects of flattening filter-free and volumetric- modulated arc therapy delivery on treatment efficiency. J Appl Clin Med Phys. 2013, 14(6):4328.

13.Prendergast BM, Fiveash JB, Popple RA, Clark GM, Thomas EM et al. Flattening filter-free linac improves treatment delivery efficiency in stereotactic body radiation therapy. J Appl Clin Med Phys. 2013, 14(3):4126.

14.Vassiliev ON, Kry SF, Chang JY, Balter PA, Titt U, Mohan R. Stereotactic radiotherapy for lung cancer using a flattening filter free Clinac. J Appl Clin Med Phys. 2009, 10(1): 2880.

15.Stieler F, Fleckenstein J, Simeonova A, Wenz F, Lohr F. Intensity modulated radiosurgery of brain metastases with flattening filter-free beams. Radiother Oncol. 2013, 109(3):448-451.

16.Chang Z, Wu Q, Adamson J, Ren L, Bowsher J et al. Commissioning and dosimetric characteristics of TrueBeam system: composite data of three TrueBeam machines. Med Phys. 2012, 39(11):6981-7018.

17.Deasy JO, Blanco AI, Clark VH.CERR: a computational environment for radiotherapy research. Med Phys. 2003, 30(5): 979-985.

18.Fowler JF. The linear-quadratic formula and progress in fractionated radiotherapy. Br J Radiol. 1989, 62(740): 679- 694.

19.Niemierko A. Reporting and analyzing dose distributions: A concept of equivalent uniform dose. Med Phys. 1997,24:103-110.

20.Niemierko A. A generalized concept of equivalent uniform dose (EUD).Med Phys. 1999, 26:1100.

21.Gay HA, Niemierko A. A free program for calculating EUDbased NTCP and TCP in external beam radiotherapy.Phys Medcia. 2007, 23(3-4):115-125.

22.Oinam AS, Singh L, Shukla A, Ghoshal S, Kapoor R et al. Dose volume histogram analysis and comparison of different radiobiological models using in-house developed software. J Med Phys. 2011, 36(4):220-229.

23.Okunieff P, Morgan D, Niemierko A, Suit HD. Radiation dose-response of human tumors. Int J Radiat Oncol Biol Phys. 1995, 32(4):1227-1237.

24.Emami B, Lyman J, Brown A, Coia L, Goitein M. Tolerance of normal tissue to therapeutic irradiation. Int J Radiat Oncol Biol Phys. 1991, 21(1):109-122.

Cite this article: Paliwal B R. The Effects of Flattened and Flattening-Filter-Free Beam on Treatment Plans. J J Rad Oncol. 2015, 2(4): 021.

Contact Us:
9600 GREAT HILLS
TRAIL # 150 W
AUSTIN, TEXAS
78759 ( TRAVIS COUNTY)
E-mail : info@jacobspublishers.com
Phone : 512-400-0398